This optical spectrometer continuously samples droplet-laden ambient air and produces size distributions in the range from 2 – 50 µm.
I am interested in this product
Close Form
Downloads
Detailed Specs
The FM-120 is a forward-scattering optical spectrometer. For accurate sizing, the FM-120 accepts and sizes only particles that pass through a region of the laser beam with uniform power. This region of the laser is called the depth of field. As particles pass through the laser beam, light scatters in all directions. The FM-120 collects forward-scattered photons within an annular cone that is approximately 3.5° to 12° from the laser beam. The collected light is then directed onto a 50/50 optical beam splitter and finally to a pair of photodetectors, referred to as the sizer and the qualifier. There is a mask in front of the qualifier detector to define the depth of field. The edge of the depth of field is defined by the points where half of the light scattered from a particle is blocked by the mask. The photodetectors then convert the photon pulses into digital pulses. If the qualifier pulse is greater than half the sizer signal, the particle is deemed within the depth of field. The particle is then sized based on the amplitude of the sizer pulse.