      Detailed Specs

      Real-time Aethalometer® method, 5 wavelength absorption analysis by measuring the rate of change of transmitted light due to continuous particle deposition on the filter. Measurement at 880 nm interpreted as concentration of Black Carbon (‘BC’). Measurement at 375 nm interpreted as Ultraviolet Particulate Matter (‘UVPM’) indicative of woodsmoke, tobacco, and biomass burning.
      Measurement Wavelengths: 880 nm, 625 nm, 528 nm, 470 nm, 375 nm

      DualSpot® Loading Compensation: Real-time analysis by measuring the rate of change in absorption of transmitted light due to the continuous collection of aerosol on filter. Simultaneous collection on two spots in parallel at different flow rates.

      Timebases: 1, 5, 10, 30, 60, or 300 seconds
      Flow Rates: Internal pump provides 50, 75, 100, 125, 150, or 170 ml/min. DualSpot® compensation not compatible with all settings.

      Measurement Range
      Per sampling location, 0-1 mg BC/m3, filter sampling location lifetime dependent on concentration and flow rate setting, decreasing proportionally with lowest wavelength optical source enabled:
      IR only mode, average 5 μg BC/m3 for 24 hours at 100 ml/min
      IR only mode, average 100 μg BC/m3 for 3 hours at 50 ml/min
      IR only mode, average 1 mg BC/m3 for 15 minutes at 50 ml/min
      Measurement Resolution: 0.001 μg BC/m3

      Limit of Detection: 30 ng BC/m3, 5 min timebase., 150 ml/min flow rate, SingleSpot™